Quantitative PET/MR imaging of lung cancer in the presence of artifacts in the MR-based attenuation correction maps

Kuttner, Samuel; Lassen, Martin Lyngby; Øen, Silje Kjærnes; Sundset, Rune; Beyer, Thomas; Eikenes, Live
Journal article, Peer reviewed
Accepted version
Thumbnail
View/Open
Kuttner (3.616Mb)
URI
http://hdl.handle.net/11250/2646891
Date
2019
Metadata
Show full item record
Collections
  • Institutt for sirkulasjon og bildediagnostikk [1049]
  • Publikasjoner fra CRIStin - NTNU [20888]
Original version
Acta Radiologica. 2019, 1-10.   10.1177/0284185119848118
Abstract
Background

Positron emission tomography (PET)/magnetic resonance (MR) imaging may become increasingly important for assessing tumor therapy response. A prerequisite for quantitative PET/MR imaging is reliable and repeatable MR-based attenuation correction (AC).

Purpose

To investigate the frequency and test–retest reproducibility of artifacts in MR-AC maps in a lung cancer patient cohort and to study the impact of artifact corrections on PET-based tumor quantification.

Material and Methods

Twenty-five lung cancer patients underwent single-day, test–retest, 18F-fluorodeoxyglucose (FDG) PET/MR imaging. The acquired MR-AC maps were inspected for truncation, susceptibility, and tissue inversion artifacts. An anatomy-based bone template and a PET-based estimation of truncated arms were employed, while susceptibility artifacts were corrected manually. We report the frequencies of artifacts and the relative difference (RD) on standardized uptake value (SUV) based quantification in PET images reconstructed with the corrected AC maps.

Results

Truncation artifacts were found in all 50 acquisitions (100%), while susceptibility and tissue inversion artifacts were observed in six (12%) and 26 (52%) of the scans, respectively. The RD in lung tumor SUV was < 5% from bone and truncation corrections, while up to 20% RD was introduced after susceptibility artifact correction, with large inconsistencies between test–retest scans.

Conclusion

The absence of bone and truncation artifacts have limited effect on the PET quantification of lung lesions. In contrast, susceptibility artifacts caused significant and inconsistent underestimations of the lung tumor SUVs, between test–retest scans. This may have clinical implications for patients undergoing serial imaging for tumor therapy response assessment.
Publisher
SAGE Publications
Journal
Acta Radiologica

Contact Us | Send Feedback

Privacy policy
DSpace software copyright © 2002-2019  DuraSpace

Service from  Unit
 

 

Browse

ArchiveCommunities & CollectionsBy Issue DateAuthorsTitlesSubjectsDocument TypesJournalsThis CollectionBy Issue DateAuthorsTitlesSubjectsDocument TypesJournals

My Account

Login

Statistics

View Usage Statistics

Contact Us | Send Feedback

Privacy policy
DSpace software copyright © 2002-2019  DuraSpace

Service from  Unit