How different iterative and filtered back projection kernels affect CT numbers and low contrast detectability
Journal article, Peer reviewed
Published version
View/ Open
Date
2016Metadata
Show full item recordCollections
Original version
Journal of computer assisted tomography. 2016, 41 (1), 75-81. 10.1097/RCT.0000000000000491Abstract
Objective: The aim of this study was to evaluate how different iterative and filtered back projection kernels affect the computed tomography (CT) numbers and low contrast detectability.
Methods: Five different scans were performed at 6 different tube potentials on the same Catphan 600 phantom using approximately the same dose level and otherwise identical settings. The scans were reconstructed using all available filtered back projection body kernels and with iterative reconstruction techniques.
Results: The CT numbers and the contrast-to-noise ratios were reported and how they are affected by the kernel choice and strength of iterative reconstruction.
Conclusions: Iterative reconstruction improved contrast-to-noise ratio in most cases, but in certain situations, it decreased it. Variations in CT numbers can be large between kernels with similar sharpness for certain densities.