Vessel-based CTA-image to spatial anatomy registration using tracked catheter position data: preclinical evaluation of in vivo accuracy
Peer reviewed, Journal article
Published version

View/ Open
Date
2024Metadata
Show full item recordCollections
Abstract
Electromagnetic tracking of endovascular instruments has the potential to substantially decrease radiation exposure of patients and personnel. In this study, we evaluated the in vivo accuracy of a vessel-based method to register preoperative computed tomography angiography (CTA) images to physical coordinates using an electromagnetically tracked guidewire. Centerlines of the aortoiliac arteries were extracted from preoperative CTA acquired from five swine. Intravascular positions were obtained from an electromagnetically tracked guidewire. An iterative-closest-point algorithm registered the position data to the preoperative image centerlines. To evaluate the registration accuracy, a guidewire was placed inside the superior mesenteric, left and right renal arteries under fluoroscopic guidance. Position data was acquired with electromagnetic tracking as the guidewire was pulled into the aorta. The resulting measured positions were compared to the corresponding ostia manually identified in the CTA images after applying the registration. The three-dimensional (3D) Euclidean distances were calculated between each corresponding ostial point, and the root mean square (RMS) was calculated for each registration. The median 3D RMS for all registrations was 4.82 mm, with an interquartile range of 3.53–6.14 mm. A vessel-based registration of CTA images to vascular anatomy is possible with acceptable accuracy and encourages further clinical testing.