Vis enkel innførsel

dc.contributor.authorAvdal, Jørgen
dc.contributor.authorLøvstakken, Lasse
dc.contributor.authorTorp, Hans
dc.contributor.authorEkroll, Ingvild Kinn
dc.date.accessioned2017-10-26T13:54:34Z
dc.date.available2017-10-26T13:54:34Z
dc.date.created2017-10-09T10:33:28Z
dc.date.issued2017
dc.identifier.issn0885-3010
dc.identifier.urihttp://hdl.handle.net/11250/2462426
dc.description.abstractMeasurement of the maximum blood flow velocity is the primary means for determining the degree of carotid stenosis using ultrasound. The current standard for estimating the maximum velocity is Pulsed Wave (PW) Doppler with manual angle correction, which is prone to error and interobserver variability. In addition, spectral broadening in the velocity spectra leads to overestimation of maximal velocities. In this work, we propose to combine two velocity estimation methods to reduce the bias and variability in maximum velocity measurements. First, the direction of the blood flow is estimated using an aliasing-resistant least squares vector Doppler technique. Then, tracking Doppler is performed on the same data, using the direction of the vector Doppler estimate as tracking direction. Simulations show that the method can estimate a maximum velocity of 2 m/s with a bias smaller than 5% for beam-to-flow angles between 20° and 75°, and that the primary source of error is inaccuracy in the flow direction estimate from vector Doppler. Simulations of complex flow in a carotid bifurcation demonstrated that the combined technique provided spectral velocity profiles corresponding well with the true maximum velocity trace, and that the bias originating from the directional estimate was within 5% for all spatial points. A healthy volunteer and a volunteer with carotid artery stenosis were imaged, showing in vivo feasibility of the method, for high velocities and with beam-to-flow angles varying throughout the cardiac cycle.nb_NO
dc.language.isoengnb_NO
dc.publisherInstitute of Electrical and Electronics Engineers (IEEE)nb_NO
dc.titleCombined 2-D Vector imaging and Tracking Doppler for improved vascular blood velocity quantificationnb_NO
dc.typeJournal articlenb_NO
dc.typePeer reviewednb_NO
dc.description.versionacceptedVersionnb_NO
dc.source.journalIEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Controlnb_NO
dc.identifier.doi10.1109/TUFFC.2017.2757600
dc.identifier.cristin1503247
dc.description.localcode© 2017 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works. Final publication available at http://ieeexplore.ieee.org/document/8052500/?reason=concurrencynb_NO
cristin.unitcode194,65,25,0
cristin.unitnameInstitutt for sirkulasjon og bildediagnostikk
cristin.ispublishedtrue
cristin.fulltextpostprint
cristin.qualitycode1


Tilhørende fil(er)

Thumbnail

Denne innførselen finnes i følgende samling(er)

Vis enkel innførsel